Test processor plug-in modules, Test processor plug-in modules 72 – MTS Model FlexTest SE User Manual

Page 72

Advertising
background image

Models FlexTest® IIm/GT/SE Controller Hardware

Test Processor Plug-in Modules

Installation

72

Test Processor Plug-in Modules

The following table provides a brief description of the plug-in modules installed
in your test processor chassis.

Test Processor Plug-In Modules

M

ODULE

F

UNCTION

Processor Modules

Standard FlexTest IIm Controllers have two processors installed in the

Model 498.22 Chassis:

The first processor (in slot 1) communicates with the external
computer.

The second processor (in slot 3) performs the digital control
functions.

A third processor will be installed (in slot 3 or 5) to handle time history
data acquisition if you purchased the time history data acquisition

option.

498.71B Global Resources III
(GRES III)

Provides the digital input and output interface to devices like external
temperature controllers and remote station controllers. It provides four

serial ports and monitors interlocks from the 497 modules.

498.65 ADDA

Provides analog-to-digital (A/D) processing and digital-to-analog (D/A)

processing. Each function is performed by a mezzanine card. The

standard ADDA module contains two A/D mezzanine cards and one D/
A mezzanine card. A fourth mezzanine card can be added as an option.

See the ADDA product manual.

493.50 ADDA II

This optional module supports up to four 8-channel A/D (Model 493.55)
or D/A (Model 493.56) mezzanine cards. This module also supports the

8-channel DSPAD, and Universal Encoder mezzanine cards.

498.70 Digital I/O

Provides the means to monitor external logic levels and contact status,
and to provide logic level outputs for external device control.

Advertising
This manual is related to the following products: