J28 Hydraulic Service Manifold Connector

Models FlexTest® IIm/GT/SE Controller Hardware

FlexTest SE Controller Connections

231

Cable specification

The J25 HPU and J51 Box In connectors are 15 pin type D male connectors.

•

15 contact type D female EMI connector.

•

Cable—24 AWG, 10 conductor with braided shield with drain wire
connected to metallized plastic backshell to the chassis.

Jumper plug required

If connector J25 HPU is not used, you must install a jumper plug to maintain the
integrity of the interlocks. Use jumper plug PN 039-713-301 or jumper pins: 1–7,
2–3–5, 6–9, 8–10–11–12, 6-9.

For an electrical diagram of J25, see

page 132

.

J28 Hydraulic Service Manifold Connector

See

“J28 Hydraulic Service Manifold Connector”

on page 134.