Accelerometer Connections

Models FlexTest® IIm/GT/SE Controller Hardware

FlexTest GT Controller Connections

119

Accelerometer Connections

An accelerometer requires the Model 493.48 Acceleration mezzanine card
(installed in a Model 493.40 I/O Carrier module) and the Model 493.78
Accelerometer Transition Board.

Each Model 493.48 Acceleration mezzanine card supports up to three
accelerometers.

The transition board accommodates two types of accelerometers: Low
impedance Voltage Mode (LIVM) type accelerometers and High Level
Signal Conditioned type accelerometers.

I/O Carrier jumper

settings

Before installing a Model 493.48 Acceleration Conditioner mezzanine card in a
Model 493.40 I/O Carrier module, specific jumpers must be set. Jumper settings
depend on the number and configuration of your system accelerometers.

Ch 1A In

Ch 2A In

Ch 3A In

m

J12 B Out

J11 B Out

Ch 1B In

Ch 2B In

Ch 3B In

493.78

Accel.

Transition

J6 I/O

J7 I/O

493.40

I/O Carrier

J3 Service

Shunt Cal

J4 I/O

J5 I/O

m

493.48 Acceleration

Conditioner

493.21B Universal

Conditioner

493.21B Universal

Conditioner

493.21B Universal

Conditioner

493.48 Acceleration

Conditioner

493.21B Universal

Conditioner

493.21B Universal

Conditioner

493.21B Universal

Conditioner

J6 I/O

J7 I/O

493.40

I/O Carrier

J3 Service

Shunt Cal

J4 I/O

J5 I/O

m

Accelerometer

Accelerometer

Accelerometer

LIVM

Accelerometer

Accelerometer

Accelerometer

High Level

Signal Conditioned

Advertising
This manual is related to the following products: