Models FlexTest® IIm/GT/SE Controller Hardware

Cabling and Using External Readout Devices

FlexTest GT Controller Connections

170

The following table provides a description of the various station signals.

Station Signals

S

IGNAL

D

ESCRIPTION

Time

Displays the time reference signal derived from the internal clock that

increments continually. This signal exists to support data acquisition by test

program applications, such as Basic TestWare and MultiPurpose TestWare. It is
not practical to monitor the time signal with an external readout device.

Rollover Time

Displays the time reference signal derived from the internal clock that resets

every hour. Like the time signal, this signal exists to support data acquisition
by test program applications. It is not practical to monitor the hourly rollover

time with external readout devices.

Output

Displays the test control signal sent to the valve driver in volts.

Command

Displays the program command signal in engineering units.

Comp. Cmd

Displays the value of the compensated command signal in engineering units.

This signal is only valid when using PVP, PVC, AIC, APC, or ALC
compensators.

Count

Displays the number of segments played out so far in the selected control

channel. When using a Station Manager meter, this signal is only available for
the Timed meter.

Error

Displays the error signal in engineering units. The error is the difference
between the command signal and the sensor feedback signal.

Abs. Error

Displays the absolute value of the error signal in engineering units.