Sensor connections, Sensor connections 88 – MTS Model FlexTest SE User Manual

Page 88

Advertising
background image

Models FlexTest® IIm/GT/SE Controller Hardware

Sensor Connections

FlexTest GT Controller Connections

88

Sensor Connections

Sensor connections require a conditioner mezzanine card be installed in the I/O
Carrier module. The following conditioners can be installed:

Note

Model 493.21B Digital Universal Conditioners are not compatible with
stand-alone FlexTest SE Controllers.

Model 493.21B Digital Universal Conditioner

Model 493.21BC Digital Universal Conditioner

Model 493.25 Digital Universal Conditioner

Model 493.47 Encoder

Model 493.48 Acceleration Conditioner

Each Carrier module can include up to four mezzanine cards. Each installed
mezzanine card is assigned a specific I/O Carrier module front panel connector
(J4–J7 I/O). These connections can be used for any type of sensor (provided the
appropriate mezzanine card is installed).

A hardware interface file (.hwi) defines each type of module (and their associated
mezzanine cards) and maps each module location for the system software. The
.hwi file and the physical locations for each type of module and associated
mezzanine cards must match.