Connecting Electrical Power

Models FlexTest® IIm/GT/SE Controller Hardware

Installation

55

Power

All equipment related to the chassis should be connected on the same fused
power circuit.

The power supply can accept single-phase voltages within 90–264 V AC at
frequencies between 47–63 Hz.

The maximum continuous power usage is approximately 300 W. The
current draw depends on the local voltage supply. A 15 amp line will be
adequate for the chassis and the computer.

The power supply automatically selects the proper voltage range and line
frequency.

The power supply is protected with an external circuit breaker in the On/Off
switch that trips at a 5 amp overload. An internal fuse in the power supply is
not user accessible or repairable.

AC power disconnect

Turn off the AC power switch. Remove the AC power cord from the unit. This
will remove all AC power from the Model 493.02 Chassis.