How to Provide Program to a Series 407 Controller

Models FlexTest® IIm/GT/SE Controller Hardware

FlexTest GT Controller Connections

155

D. On the 407 Controller, clear the hydraulic Interlock:

Press the Alt Function button.

–Press the Reset button (“9” on the key pad). The Interlock indicator

on the front panel should extinguish.

E.

On the 407 Controller, apply hydraulic pressure with the Hydraulic
Pressure
buttons.

F.

On the 407 Controller, press Run. This step readies the 407 Controller
to receive program from the 793 Controller.

G.

On the Station Manager application’s Station Controls panel, click
Reset to clear the Program interlock.

H. On the Station Manager application’s Station Controls panel, click the

(Program Run). This step sends program from the 793 Controller

to the 407 Controller.