MTS Model FlexTest SE User Manual

Page 206

Advertising
background image

Models FlexTest® IIm/GT/SE Controller Hardware

How to Program an External Controller

FlexTest IIm Controller Connections

206

Note

Analog I/O connectors are located on the rear panel of the Series 498
test processor chassis.

1. Connect a 493.82 Analog Out BNC connector to the external controller

program (analog) input.

2. If you will monitor sensor feedback for data acquisition or control feedback,

connect the external conditioner output to a 493.83 Filtered Analog Input
BNC connector.

For conditioner output locations on various MTS devices, see

“Sample

conditioner output connections”

on page 207.

Transition Module Connections

Transition modules on the rear panel of the
Series 498 test processor chassis are used for
analog inputs and outputs.

493.83 Filtered Analog Input
This module includes 8 BNC
connectors that provide filtered
analog inputs to the Model
493.55 A/D mezzanine cards.
Eight analog inputs are
available with the each A/D
mezzanine card installed on
the Model 493.50 ADDA II
module.

493.82 Analog Out
This module includes 8 BNC
connectors that provide analog
outputs from a Model 493.56 D/A
mezzanine card.
Eight analog outputs are available
with each D/A mezzanine card
installed on the Model 493.50
ADDA II module.

Advertising