Or 8-wire sensor cable – MTS Model FlexTest SE User Manual

Page 95

Advertising
background image

Digital Universal Conditioner (DUC BC) Switches

Models FlexTest® IIm/GT/SE Controller Hardware

FlexTest GT Controller Connections

95

4- or 8-Wire

sensor cable

Switches S1-1, S1-2, S1-3, and S1-4 establish the correct shunt calibration
connections for 4-wire and 8-wire sensor cables. The length of the sensor cable
determines which type of cable is used. Short cables are typically 4-wire cables
while long cables are typically 8-wire cables.

Switches S1-1 and S1-2 select a 4- or 8-wire transducer for positive
feedback.

Switches S1-3 and S1-4 select a 4- or 8-wire transducer for negative
feedback.

The setting of jumper W8 (single ended excitation enable) may affect your
choices for these switches.

Note

+/– excitation (–excitation is standard).

Sensor Cable Settings (4/8-Wire)

REQUIRED

CONNECTION

SWITCH

SETTINGS

-EX to -FB shunt in 4-wire
configuration

DEFAULT

S1-1 OFF

S1-2 OFF

S1-3 ON

S1-4 OFF

-EX to +FB shunt in 4-wire
configuration

S1-1 ON

S1-2 OFF

S1-3 OFF

S1-4 OFF

-EX to -FBR shunt in 8-wire
configuration

S1-1 OFF

S1-2 OFF

S1-3 OFF

S1-4 ON

-EX to +FBR shunt in 8-wire
configuration

S1-1 OFF

S1-2 OFF

S1-3 ON

S1-4 OFF