Analog inputs – MTS Model FlexTest SE User Manual

Page 115

Advertising
background image

Analog I/O Connections

Models FlexTest® IIm/GT/SE Controller Hardware

FlexTest GT Controller Connections

115

Analog inputs

The analog-to-digital mezzanine cards accommodate up to six analog input
signals. Each A/D input signal must be within ±10 volts.

Use pin 3 or 6 to prevent problems with floating grounds between devices.

Analog inputs can be connected at the BNC connectors of a Model 493.75
Analog In transition module and the outputs from this module are connected
to the Model 493.40 I/O Carrier module.

Analog inputs can be connected at the BNC connectors of a Model 493.77
Filtered Analog Input transition module and the outputs from this module
are connected to the Model 493.40 I/O Carrier module.

+

-

+

-

+

-

Input 1

Analog

Ground

Input 2

Analog

Ground

Input 3

Input 4

Input 5

Input 6

+

-

+

-

+

-

+

-

1
2
3
4
5
6
7
8
9

10

11

12
13
14
15

J4 - J7

To

A/D

From

A/D

Advertising