MTS Model FlexTest SE User Manual

Page 97

Advertising
background image

Full-Range Digital Universal Conditioner (FRDUC)

Models FlexTest® IIm/GT/SE Controller Hardware

FlexTest GT Controller Connections

97

Full-Range Digital Universal Conditioner (FRDUC) Jumpers

The following jumpers are for the Model 493.25 Full-Range Digital Universal
Conditioner. The jumpers are labeled W instead of X; the W jumpers are solder
pads. The following jumper descriptions are from the 493.xx Module Setup
configuration drawing (PN 049-336-301).

Active guard

Jumper W1 enables or disables an active guard for the sensor cable. A passive
guard drive is the default.

4- or 8-wire

sensor cable

Jumpers W2 and W3 establish the correct shunt calibration connections for 4-
wire and 8-wire sensor cables. The length of the sensor cable determines which
type of cable is used. Short cables are typically 4-wire cables while long cables
are typically 8-wire cables.

Jumper W2 selects a 4- or 8-wire transducer for positive feedback.

Jumper W3 selects a 4- or 8-wire transducer for negative feedback.

The setting of jumper W8 (single ended excitation enable) may affect your
choices for these jumpers (W2 and W3).

Note

+/– excitation (–excitation is standard).

W1

W1

Enable

Disable

Default

(Short to Ground)

(Cap to Ground)

Default

W2

W3

W2

W3

1

2

3

W2
W3

W2
W3

–EX to –FB shunt
in 4-wire config

-EX to +FB shunt
in 4-wire config

–EX to –FBR shunt
in 8-wire config

-EX to +FBR shunt
in 8-wire config