MTS Model FlexTest SE User Manual

Page 194

Advertising
background image

Models FlexTest® IIm/GT/SE Controller Hardware

Interlock Jumper Plugs

FlexTest IIm Controller Connections

194

The interlock connectors require jumpers under the following conditions:

If you do not use the external hydraulic/program interlocks on any of the D-
15S connectors (J3A Sta 1, J3B Sta 2, J3C Sta 3, and J3D Sta 4). Use an
external jumper plug (PN 052-750-501) on each connector not used.

If you use the system digital I/O as a stand-alone unit. That is, if you do not
connect the System Digital I/O Module to a 497.xx Chassis, you must use an
external jumper plug (PN 050-774-701) on the J5 497 Intlk connector.

This shows how to connect an external switch or set of relay contacts to one of
the station interlocks:

J3A-J3D

External Switch

or

Relay Contact

External Device

+12 VDC

x9—x11

x1—x4

Opto-

Coupler

1

9

7