Models FlexTest® IIm/GT/SE Controller Hardware

J44 Run/Stop

FlexTest GT Controller Connections

138

J44 Run/Stop

Run/Stop status

Connector J44 Run/Stop provides the run/stop status of the controller to external
devices.

Two form C contacts provide the run/stop status.

The contacts are rated 1.0 A at 30 V (AC or DC).

Interlock status

Connector J44 Run/Stop also provides opto-isolator outputs that indicate the
interlock status of each station to an external device. These interlock status
outputs are normally on and will turn off when an interlock occurs. For detailed
circuit drawings, see

“Digital outputs”

on page 143.

Run/Stop 1

+

-

Interlock Status

Output

1

2

3

4

5

6

7

8

Run/Stop 0

J44

To

External Devices

From

Chassis