Proper grounding is required for safe operation.

It is also required to meet EMC emission and susceptibility requirements.

Power

All equipment related to the chassis should be connected on the same fused
power circuit.

The power supply can accept single-phase voltages within 90–264 V AC at
frequencies between 47–63 Hz.

The maximum continuous power usage is approximately 1000 W. The
current draw depends on the local voltage supply. A 15 amp line should be
adequate for the chassis and the computer.

The power supply automatically selects the proper voltage range and line
frequency.

The power supply is protected with an external circuit breaker in the On/Off
switch that trips at a 10 amp overload. An internal fuse in the power supply
is not user accessible or repairable.

An outlet strip is supplied with the floor-standing chassis.

The computer components may be plugged directly into the outlet strip of a
vertical console or a floor-standing console.

AC power disconnect

Turn off the AC power switch. Remove the AC power cord from the unit. This
will remove all AC power from the Model 493.10 Chassis.