MTS Model FlexTest SE User Manual

Page 201

Advertising
background image

Encoder/Temposonics Connections

Models FlexTest® IIm/GT/SE Controller Hardware

FlexTest IIm Controller Connections

201

Jumper configurations

The Series 498 transition module can be configured for a Temposonics sensor, a
5 V DC encoder, a 12 V DC encoder, a 24 V DC encoder. The following is a list
of the configuration drawings:

049-035-301 Absolute Encoder

049-035-401 +5 Volt Encoder

049-035-501 +12 Volt Encoder

049-035-601 +24 Volt Encoder

049-035-701 Temposonics II

053-070-201 Temposonics III

ADDA mezzanine

cards

The ADDA module requires special mezzanine cards to support encoders and
Temposonics sensors.

Mezzanine cards for encoders and Temposonics sensors support four input
channels.

Mezzanine cards for multiturn encoders (Heidenhain 417/425) and Temposonics
III position sensors support two channels.)

The encoder mezzanine card counts pulses from the encoder to determine
the displacement of an actuator.

The Temposonics mezzanine card interprets the pulses from the
Temposonics sensor to determine the displacement of an actuator.

Optional ADDA II

configuration

For the optional ADDA II configuration (using the Model 493.50 ADDA II
module), each encoder requires installation of a Model 493.59 Universal Encoder
mezzanine card on the Model 493.50 ADDA II module.

Advertising