Models FlexTest® IIm/GT/SE Controller Hardware

FlexTest IIm Controller Connections

211

Task 1

Cable the controllers.

Cable the external programmer analog out connector to a Series 498 Input Filter
connector. Analog I/O connectors are located on the rear of the test processor
chassis.

Note

If you cable the hardware before you create a station configuration file,
note the connector you use on the FlexTest IIm rear console.

Task 2

Allocate external command resources in Station

Builder.

Using the Station Builder application, allocate controller resources to receive
programming from an external controller.

For more information, see “Enabling External Command Inputs” in the MTS 793
Control Software
manual.

J11
(CH 1-4)

J12
(CH 5-8)

498

INPUT

FILTER

CH 1

CH 2

CH 3

CH 4

CH 5

CH 6

CH 7

CH 8

498 Input Filter
This module includes 8 BNC
connectors that provide filtered
analog inputs to the 498.65
ADDA module.
Four analog inputs are available
with the standard ADDA module.
The other four inputs require a
second ADDA module.