Digital universal conditioner (duc bc) switches, Digital universal conditioner (duc bc) switches 94 – MTS Model FlexTest SE User Manual

Page 94

Advertising
background image

Models FlexTest® IIm/GT/SE Controller Hardware

Digital Universal Conditioner (DUC BC) Switches

FlexTest GT Controller Connections

94

Excitation

Jumpers W5 and W7 select the local (on board) or remote (through a cable)
excitation sensor connection.

The local selection is used with 4-wire transducer connections.

The remote selection is used with 8-wire transducer connections.

Jumper W7 configures the + excitation signal.

Jumper W5 configures the – excitation signal.

Both jumpers need to be set for the same configuration.

Single ended

excitation

Jumper W8 connects -EX to ground for situations requiring single- ended
excitation.

Digital Universal Conditioner (DUC BC) Switches

The following switch settings are for the Model 493.21BC Digital Universal
Conditioner. The switches are labeled S1 1-8 and replace the former solder pads.
The following descriptions are from the 493.XX Module Setup configuration
drawing (PN 049-336-301).

Active guard

Switch S1-8 enables or disables the guard connection to ground (GND) for the
sensor cable.

W7
W5

W7
W5

Local

Remote

Default

Default

–EX = –EX

–EX = GND

W8

W8

Enable

Disable

Default

Short to Ground

Cap to Ground

S1-8 ON

S1-8 OFF