Models FlexTest® IIm/GT/SE Controller Hardware

J49 Auxiliary Power

FlexTest GT Controller Connections

140

J49 Auxiliary Power

Connector J49 Aux Pwr provides +5 V DC, ±15 V DC, and 24 V DC from the
chassis internal power supply to drive external high level conditioners, proximity
switches, solenoids, and so forth.

The outputs are fused at 0.75 A to protect the power supply from an external
short.

The fuses can be reset by shutting off power and waiting a few minutes (the
fuses automatically reset when cooled) then reapplying power.

Cable specification

9-contact type D male EMI connector

Backshell–EMI metallized plastic

Cable–shielded twisted pairs (22 AWG minimum) with drain wire(s)
connected to the metallized backshell at the chassis.

F4

F9

F7

F7

1

6

2

7

3

9

8

5

4

J49

To

External Devices

From

Chassis

-15 V DC

+24 V DC

24V

Common

+5 V DC

5V

Ground

+15 Vdc

Analog

Ground