Models FlexTest® IIm/GT/SE Controller Hardware

Connecting Interlock Signals to 407 Controllers

FlexTest GT Controller Connections

156

Connecting Interlock Signals to 407 Controllers

The Model 793 Controller monitors digital I/O interlock signals through the
Model 493.74 HSM transition module’s J43A and J43B connectors. Use these
connectors when connecting to Model 407 Controllers.

Model 493.74 HSM transition module connectors are as shown in the figure that
follows.

Interlock cabling to Model 407 Controllers depends on your system’s
configuration. There are three possible configurations:

Single station—one Model 407 Controller runs a single test program.

Multiple stations—two or more independent Model 407 Controllers each
run different test programs.

Master-dependent station—one master Model 407 Controller daisy-chained
with other dependent Model 407 Controllers together run a single test
program.

Single 407 Controller

interlock connections

The following figure shows how to cable interlocks for a single station Model
407 Controller to the Model 493.10 Chassis (via its Model 493.74 HSM
transition module). Use the specified interlock cable (PN 056-455-0xx) to make
these connections.

Run/Stop,

ActiveHigh

Hydraulic

Interlock,

ActiveLow To

407

Controller

To

493.74

HSM

Transition

Interlock Cable PN 056-455-0XX

Digital

I/O Connector

Jumper

Jumper

Jumper