J9 Interlock OUT

Models FlexTest® IIm/GT/SE Controller Hardware

FlexTest GT Controller Connections

173

Cable specifications

Connector: 25-contact, type D, male, EMI

Backshell: EMI, metalized plastic

Twisted pairs with overall braided shield. Braided shield is terminated to
conductive backshell at the chassis.

Jumper plug

Use MTS jumper plug PN 100-079-233.

J9 Interlock OUT

Connector J9 Interlock OUT provides four station interlock outputs that can be
connected to a multiple MTS controller box interlock chain.

Electrical

specifications

Normally, Open = Interlock, 1 A maximum @ 30 V DC/AC maximum

Cable specification

Connector: 9-contact, type D, male, EMI

Backshell: EMI metalized plastic

Twisted pairs with overall braided shield. Braided shield is terminated to
conductive backshell at the chassis.

+

-

Interlock Status

Output 2

+

-

Interlock Status

Output 1

+

-

Interlock Status

Output 3

+

-

Interlock Status

Output 4

1

6

2

7

3

8

4

9

J9

To

External Devices

From

Chassis

Advertising