Interlock jumper plugs, Interlock jumper plugs 193 – MTS Model FlexTest SE User Manual

Page 193

Advertising
background image

Interlock Jumper Plugs

Models FlexTest® IIm/GT/SE Controller Hardware

FlexTest IIm Controller Connections

193

Interlock Jumper Plugs

Several jumper plugs must be installed to maintain the integrity of the interlock
chain. Each jumper plug represents a place where you can add a custom
connection to the interlock chain.

Model 497.05

Hydraulic Control

Panel

Any interlocks connected to the hydraulic control panel will generate a hydraulic
interlock which shuts down the hydraulic power supply. This type of interlock
shuts down all test stations.

•

Connector J52 requires a jumper plug to close the interlock chain.

•

If J26 is not connected, it needs a jumper plug.

•

Connectors J15A–J17C (Remote Interlock Inputs) each require a jumper
plug unless it is being used. Jumper both pins of each connector.

Series 498 System

Digital I/O

Any interlock connected to the Series 498 System Digital I/O transition module
can generate a station interlock to disable pressure (the HSM) to that station.

The Series 498 System Digital I/O transition module supports:

•

Four stations of Hydraulics On digital inputs

•

Four stations of Hydraulics Hi digital inputs

•

Four stations of Hydraulics Enables

•

Four stations of Remote Stop/Run outputs

•

Four stations of program interlocks

•

Four stations of auxiliary interlocks

•

Four stations of 497 interlock inputs

•

Four stations of 497 interlock outputs

•

Four stations of auxiliary interlock inputs

•

Four stations of auxiliary interlock outputs

•

Two spare digital inputs

•

Two spare digital outputs