Interlock Jumper Plugs

Models FlexTest® IIm/GT/SE Controller Hardware

FlexTest IIm Controller Connections

193

Interlock Jumper Plugs

Several jumper plugs must be installed to maintain the integrity of the interlock
chain. Each jumper plug represents a place where you can add a custom
connection to the interlock chain.

Model 497.05

Hydraulic Control

Panel

Any interlocks connected to the hydraulic control panel will generate a hydraulic
interlock which shuts down the hydraulic power supply. This type of interlock
shuts down all test stations.

Connector J52 requires a jumper plug to close the interlock chain.

If J26 is not connected, it needs a jumper plug.

Connectors J15AJ17C (Remote Interlock Inputs) each require a jumper
plug unless it is being used. Jumper both pins of each connector.

Series 498 System

Digital I/O

Any interlock connected to the Series 498 System Digital I/O transition module
can generate a station interlock to disable pressure (the HSM) to that station.

The Series 498 System Digital I/O transition module supports:

Four stations of Hydraulics On digital inputs

Four stations of Hydraulics Hi digital inputs

Four stations of Hydraulics Enables

Four stations of Remote Stop/Run outputs

Four stations of program interlocks

Four stations of auxiliary interlocks

Four stations of 497 interlock inputs

Four stations of 497 interlock outputs

Four stations of auxiliary interlock inputs

Four stations of auxiliary interlock outputs

Two spare digital inputs

Two spare digital outputs