Models FlexTest® IIm/GT/SE Controller Hardware

Shared HSM (no HPU)

Hydraulic Configurations

248

For this multiple controller configuration:

The HSM and HPU are connected to one controller only.

Select HPU/HSM for Power Type on the controller to which the HSM and
HPU is connected.

Path: Config > Edit Hardware > Power Options > Power Type > HPU/
HSM

Select Slave HSM for Power Type on controllers connected to the master
controller.

Shared HSM (no HPU)

For this multiple controller configuration:

An HSM is connected to one controller only.

Select HSM Only for Power Type on the controller to which the HSM is
connected.

Path: Config > Edit Hardware > Power Options > Power Type > HSM
Only

Select Slave HSM for Power Type on controllers connected to the master
controller.

FlexTest SE

Emergency Stop

7

8

9
6

5

4

3

2

1

.

0

+/-

enter

cancel

FG

Setup

Status

Scope

Meters

Tuning

Config

Limits

Log

Menu

?

A

B

Recall

Enabled

reset

Rewind

Run

Hold

Stop

Power

Program

Hydraulic

Interlock

Program

Interlock

HPU

HSM

Off

Low

High

Off

Low

High

Navigate

Monitor 1

Monitor 2

FlexTest SE

Emergency Stop

7

8

9
6

5

4

3

2

1

.

0

+/-

enter

cancel

FG

Setup

Status

Scope

Meters

Tuning

Config

Limits

Log

Menu

?

A

B

Recall

Enabled

reset

Rewind

Run

Hold

Stop

Power

Program

Hydraulic

Interlock

Program

Interlock

HPU

HSM

Off

Low

High

Off

Low

High

Navigate

Monitor 1

Monitor 2

Power Type = HSM Only

Power Type = Slave HSM

Interlock/Hydraulics

056-031-6xx

HSM

056-126-4xx

J28 HSM

Advertising
This manual is related to the following products: