Optional station configurations, Station configuration, Appendix d – MTS Model FlexTest SE User Manual

Page 255: Optional station configurations 255, Station configuration 255

Advertising
background image

6-Station Configuration

Models FlexTest® IIm/GT/SE Controller Hardware

Optional Station Configurations

255

Appendix D

Optional Station Configurations

This section describes how to configure your Model 493.10 Chassis (FlexTest GT
Controllers) to support an optional six or eight stations.

When configuring your system for a six or eight stations you must consider the
following:

Ensure that the .hwi file is correctly set for the desired multi-station
configuration, especially the interlock and HSM board settings.

Power to each HSM is limited.

Cross-head interlocks with solenoid power are not supported.

Remote Station Control (RSC) is not supported.

AC input power must be at least 115 V AC.

Auxiliary power out of J49 on the Model 493.74 HSM Transition board is
not supported.

6-Station Configuration

The 6-station configuration can provide either six or eight channels of control.

A typical 6-channel/6-station configuration requires 12 Digital Universal
Conditioners and 6 two-stage valve drivers.

A typical 8-channel/6-station configuration requires 16 Digital Universal
Conditioners and 8 two-stage valve drivers.

HSM power limits

HSM power current is limited to 1.5 A per HSM.

Interlocks

For 6-station configurations the .hwi file must contain the line
INTERLOCKS=6. This line must be a discrete entry, not part of any other .hwi
section.

Cross-head interlocks with solenoid power are not supported for 6-station
configurations.

RSC

Remote Station Control (RSC) is not supported for 6-station configurations.