Vmebus modules, Vmebus modules 50 – MTS Model FlexTest SE User Manual

Page 50

Advertising
background image

Models FlexTest® IIm/GT/SE Controller Hardware

VMEbus Modules

Installation

50

VMEbus Modules

The VMEbus compatible modules should be installed in chassis slots according
to the following guidelines.

The processor module is installed in slot 1

The I/O carrier module is installed in slot 2.

The System I/O module is installed in slots 4 and 5.

VMEbus Modules

S

LOT

M

ODULE

N

AME

F

UNCTION

Slot 1

Processor

Provides PIDF processing for the Controller and
management of the other plug-in modules in the standalone
configuration. Provides an interface between the Controller
and an external computer in the automated configuration.

Slot 2

Model 493. 40 I/O Carrier

Supports up to four mezzanine cards that can be used for
sensor conditioning, driving a servovalve, and in the
automated configuration, providing A/D and/or D/A channels.

Slot 3

Blank in standard
configurations.

May be used for additional
Model 493.40 I/O Carrier
(factory installed option for
automated systems)

Additional carrier chassis that can house up to four
mezzanine cards for expanded sensor conditioning and A/D
and D/A channels

Note

This is not a standard option. It is available only for
automated systems and must be configured at the
factory. For more information, contact MTS.

Slots 4, 5

Model 493.42 System I/O

Provides box-in and box-out connections for daisy chaining
multiple Controllers, analog and digital I/O, E-Stop, and HSM
and HPU control.

5

4

3

2

1

TYPE E PC-MIP

TYPE E PC-MIP

RUN 6TS BFL

ABT

RST

DEBUG

PCI MEZZANNE CARD

Placement of

VMEbus modules

in the chassis

Advertising