Model 497.01 analog chassis, About analog chassis slots, Model 497.01 analog chassis 63 – MTS Model FlexTest SE User Manual

Page 63: About analog chassis slots 63

Advertising
background image

Model 497.01 Analog Chassis

Models FlexTest® IIm/GT/SE Controller Hardware

Installation

63

Model 497.01 Analog Chassis

About Analog Chassis Slots

Slot pairing

Chassis slot assignments are grouped in pairs. If you have a conditioner in a slot,
then you cannot have a valve driver installed in the slot paired with it. Slots are
paired as follows:

Slot assignments

The analog chassis contains 18 bus-board slots for plug-in modules. Each slot is
reserved for a particular type of plug-in module.

Slot A

Reserved for the Model 497.34 Multistation Interlock Module.

Slot B

Reserved for the Model 497.36 Communications Module.

Slots 1–4

These slots are reserved for up to four AC conditioners.

Slots 5–8

These slots are reserved for up to four DC conditioners.

Slots 9–12

These slots are reserved for options (not used with standard systems).

Slots 13–16

These slots are reserved for up to four valve drivers.

Slots 1–2
Slots 3–4
Slots 5–6
Slots 7–8

Slots 9–10
Slots 11–12
Slots 13–14
Slots 15–16