Model 497.05 Hydraulic Control Panel

Models FlexTest® IIm/GT/SE Controller Hardware

Installation

71

Model 497.05 Hydraulic Control Panel

The rear panel connectors of the hydraulic control panel are cabled to the
hydraulic components of your system. The connectors also interface with the
Model 497.01 Analog Chassis and Model 498.22 Test Processor chassis.

Hydraulic Control Panel

J1

The 497.05 contains the power supply for a Model 497.01 Analog Chassis.
Connect your 497.01 Analog Chassis to J1.

J8

This connector communicates hydraulic status with the Model 498.22 Test
Processor chassis.

J9

This connector is not used in a standard FlexTest IIm system.

J15A–J17D

Each HSM channel has a set of three user defined interlock inputs (Interlock 1,
Interlock 2, and Shutdown). These inputs can be configured to monitor contact
closure or to accept logic level inputs.

J25

This connector controls the hydraulic power supply (HPS).

J26

This connector accommodates a remote Emergency Stop switch.

J28A–J28D

These connectors control up to four hydraulic service manifolds (HSMs).

J41 and J42

These connectors create a serial communications loop between all of the chassis
in your system.

J51 and J52

These connectors create an interlock communications loop with the other chassis
in your system.

I

O

J41

Com In

J8

Hyd Enab

J15C

J16C

J17C

J15B

J16B

J17B

J15A

J16A

J17A

J1

497.01 Power

J26

Remote E-Stop

J25

HPS

J52

Intlk Out

J51

Intlk In

J42

Com Out

J9

Mstr/Slv

J28D

J15D

J16D

J17D

J28C

J28B

J28A