Model 493.42 system i/o connections, J25 hydraulic power unit connection, Model 493.42 system i/o connections 230 – MTS Model FlexTest SE User Manual

Page 230: J25 hydraulic power unit connection 230

Advertising
background image

Models FlexTest® IIm/GT/SE Controller Hardware

Model 493.42 System I/O Connections

FlexTest SE Controller Connections

230

Model 493.42 System I/O Connections

The Model 493.42 System I/O module is a dual-wide VME-based slave board
used in the Model 493.02 Chassis.

Provides box-in and box-out, analog and digital I/O, E-Stop, and HSM and
HPU control.

Contains three general purpose digital input channels and three general
purpose digital output channels.

Interfaces the controller with a hydraulic power unit.

Interfaces the controller with a hydraulic service manifold and other
devices.

Provides one BNC channel for analog output signals. The output signals are
within ±10 V DC.

Provides one BNC channel of analog input. The input signals must be
within ±10 V DC.

J25 Hydraulic Power Unit Connection

Connector J25 HPU controls the hydraulic power unit (HPU).

The connector provides 24 V logic signals to control the HPU.

The connector may be connected directly to MTS Series 505 HPUs, other
MTS HPUs via the Model 493.07 HPU Converter Box, and
24 V PLC compliant HPUs.

Note

For more information, see

“Hydraulic Configurations”

on page 243.

Control voltages for hydraulic power units vary between models.

The HPU interface between the Model 493.42 System I/O board and an HPU
is 24 volt logic signals. Connecting J25 to a non-compliant HPU can damage
the module.

DO NOT connect 24 V DC relay circuitry or 115 V AC circuitry to connector J25 on
the Model 493.42 System I/O board.

CAUTION

Advertising
This manual is related to the following products: