VT MOBILE
Operators Manual

B-2

High-Pressure Port

Maximum Applied Pressure:

125 psi

Operating Pressure:

- 2 psi to 100 psi

Span Accuracy:

± 2 % of reading or ± 0.2 psig, whichever is greater

Frequency Response:

> 10 Hz

Resolution:

0.1 psi

Sample Rate:

100 Hz

Fittings:

Single port, Luer lock, stainless steel

Note: Apply no fluid to the port.

Airway Pressure

Maximum Applied Pressure:

5 psi

Operating Pressure:

- 20 cmH

2

O to 120 cmH

2

O

Span Accuracy:

± 2 % of reading or ± 2 cmH

2

O, whichever is greater

Frequency Response:

> 25 Hz or t

10-90

< 40 ms

Resolution:

0.1 cmH

2

O

Sample Rate:

100 Hz

Fittings:

Internally connected to flow sensor pressure lines