Checking ventilator parameters, Introduction, Chapter 4 – Fluke Biomedical VT Mobile User Manual

Page 45

Advertising
background image

4-1

Chapter 4

Checking Ventilator Parameters

Introduction

To monitor ventilator parameters, the Analyzer must first determine when a breath occurs

and delineate the various phases of the breath. For example, the breath detection

algorithm determines the beginning and ending of the integration periods for calculating

tidal volumes.
The breath detection algorithm is a state machine. The flow waveform is the input to the

state machine. The Analyzer bases state changes on flow waveform crossings of

predefined thresholds. The breath detection threshold can be set to a high flow rate to

reject noise near zero flow.
You can use the High-Flow or Low-Flow Sensor to sample ventilator parameters.

Generally, the Low-Flow Sensor provides better accuracy at low flow rates (less than 25

lpm). Also, review “Making a Measurement More Accurate” in Chapter 3 for a

discussion of the uses of the Oxygen Sensor, Temperature and Relative Humidity Sensor,

and barometric pressure readings in enhancing measurement accuracy. Chapter 3 also

discusses proper warm-up and zeroing procedures to compensate for sensor drift.

Checking Ventilator Parameters

Refer to Figure 4-1 when you are connecting the Analyzer to a ventilator and test lung.
1. Press R and follow on-screen instructions.
2. At the Analyzer, connect the High-Flow Sensor to the Flow/Low-Pressure Port.
3. Connect the Oxygen Sensor to the upper port on the right side of the Analyzer.
4. If available, connect the optional Temperature/RH Sensor to the lower port on the

right side of the Analyzer.

Advertising