Replacing the oxygen sensor, Calibrating the oxygen sensor – Fluke Biomedical VT Mobile User Manual

Page 63

Advertising
background image

Maintaining and Troubleshooting the VT MOBILE

Replacing the Oxygen SensorReplacing the Oxygen Sensor

7

7-3

Oxygen Sensor

Source

edg20f.eps

Figure 7-2. Oxygen Sensor Calibration

Replacing the Oxygen Sensor

The Analyzer uses a galvanic oxygen sensor warranted for 12 months. Replace the

oxygen sensor cell when you cannot successfully complete the oxygen sensor

verification. The oxygen sensor may operate for longer than 12 months depending on use.

To replace the oxygen sensor, unscrew the cable and pull the sensor cell out of its T

connector.
Follow the oxygen sensor manufacturer's unpacking and installation instructions. You

may need to expose oxygen sensors to room air for 24 hours prior to installation or

calibration.

Calibrating the Oxygen Sensor

Calibrating the Oxygen Sensor improves pressure compensation for barometric and

airway pressure changes.
1. Press P to display the Oxygen measurement screen.
2. Press G to access the Oxygen calibration screen.
3. Connect the High-Flow Sensor and then apply dry air (20.9 % oxygen) at 10 lpm to

either side of the sensor.

4. Press F twice to access the Calibration screen. Wait for the two-minute timer to

count down.

5. Apply 100 % oxygen at 10 lpm.
6. Press G twice and then press F once to access the Calibration screen.
7. After the two-minute interval, the Oxygen measurement screen appears and the

calibration is complete.

Advertising