Measuring pressure, Introduction, Chapter 5 – Fluke Biomedical VT Mobile User Manual

Page 49

Advertising
background image

5-1

Chapter 5

Measuring Pressure

Introduction

The Analyzer detects a Low-Pressure Adapter connection to the Flow/Low-Pressure Port.

With the Low-Pressure Adapter disconnected, pressing L under WAVEFORM

SCREENS or OTHER MEASUREMENTS toggles between high pressure and airway

pressure. With the Low-Pressure Adapter connected, pressing L successively selects

high pressure, low pressure, or airway pressure.
With a pressure connection, the measurement is equivalent to a gauge measurement. For

gauge pressures, the measurement is relative to the ambient (barometric) pressure.
Do not use either the High-Pressure or Low-Pressure Adapter for fluid measurement. If

fluid enters into the pressure measurements connections of the Analyzer, evacuate the

fluid before storage and before you make any pressure measurements. You can evacuate

the fluid by connecting a 100 cc syringe directly to the pressure measurement connection

and then pulling back the syringe to create a vacuum.
Review “Making a Measurement More Accurate” in Chapter 3 for a discussion of the

uses of the Oxygen Sensor, Temperature and Relative Humidity Sensor, and barometric

pressure readings in enhancing measurement accuracy. Chapter 3 also discusses proper

warm-up and zeroing procedures to compensate for sensor drift.

Advertising