VT for Windows uses a simple means to annotate data files. Each of the annotations is

marked as an "event" in the data file. When you play data files back, the annotated text

and the time of the event appear on the screen.

Events

VT for Windows supports six different types of events:

1. Start of file
2. Beginning of a recording
3. Time stamp
4. Annotation
5. End of file
6. Automatic event

Each event is marked in the .vtp data file with a time stamp. For the beginning of a

recording and for an annotated event, a dialog box appears with a space provided for you

to enter the text annotation.

Zeroing

Periodically, it is a good idea to zero the signals on the Analyzer to null any offset errors

in the flow and pressure sensors. You can set the Analyzer to zero flow and airway

pressure. This nulls offset errors in the pressure transducers. Do not use this option if

there is flow in a flow port. See Figure A-7.
VT for Windows offers two different ways to perform a zero:


1. Select Zero All from the Zeroing menu.

edg72s.bmp

Figure A-31. Zeroing Menu

2. Select Zero All with the Zeroing button on the toolbar.
The Oxygen selection allows you to zero out any offset for drift when air (21% oxygen)

passes through one of the flow sensors. You can then calibrate the Oxygen Sensor by

applying a flow of 100% oxygen.