Full test screen, Conducting a trend test, Overview – Fluke Biomedical VT Mobile User Manual

Page 77: Starting a trend test, Running a trend test

Advertising
background image

Appendices

VT for Windows

A

A-11

Full Test Screen

The Full Test screen (Figure A-14) displays the calculated parameters for the last breath.

These parameters update only when a new breath occurs. A "*****" value means that

value is invalid or the software could not calculate the value. Depending on your breath-

detect mode, some parameters may not be calculable.

edg60s.bmp

Figure A-14. Full Test Screen

Conducting a Trend Test

Overview

The Trend test monitors a single-breath parameter over an extended period. Each time a

breath occurs, the software compares the chosen breath parameter to an initial value. If it

exceeds the percent limit given for the test, a reported incident occurs. Following the test,

you can print a report detailing all the incidents that occurred during the test.

Starting a Trend Test

Initiate a Trend test by selecting Trend Test from either the Test Settings menu or the Full

Test screen. The Trend Test setup dialog box opens, allowing you to choose the

parameter on which to run the test, the percent limit for the test, and minimum and

maximum limits for the test. Click the OK button to start the test. See Figure A-15.

Running a Trend Test

During a Trend test, you see several values every time a breath occurs. Refer to Figure A-

16.

Test Parameter

This is the parameter chosen for running this test. This value does not change during
the test.

Initial Value

This is the initial value for the test parameter. This value does not change during the
test.

Current Value

This is the current value of the chosen test parameter. Every time a breath occurs, the
software updates and compares the current value against the initial value of the test
parameter.

Advertising