Measuring low pressure, Vt mobile, Operators manual – Fluke Biomedical VT Mobile User Manual

Page 50: Figure 5-1. measuring low pressure, Low pressure adapter

Advertising
background image

VT MOBILE

Operators Manual

5-2

Measuring Low Pressure

Measure low pressure in the range: -20 to 120 cmH

2

O. Never exceed an absolute

maximum of 5 psi (0.345 bar).
Refer to Figure 5-1 while following these steps:
1. Press R and follow on-screen instructions.
2. At the Analyzer, connect the Low-Pressure Adapter to the Flow/Low-Pressure Port.
3. Press L in the OTHER MEASUREMENTS or WAVEFORM SCREENS section.
4. Verify that the screen displays Low Pressure (Plo) or LPress. If the High-Pressure

Adapter is also connected, you may need to press L a second time.

5. Connect the Low-Pressure Adapter to the source.
6. Apply pressure.
7. Successively press G (Units) to select a pressure unit in the sequence: PSI Æ kPa

Æ Bar Æ mBar Æ ATM Æ inH2O Æ InHg Æ cmH2O Æ mmHg.

If the waveform is too small or too large, press F (Rescale) to view the waveform

within optimized limits.

F1

F2

Back

ENTER

Setup

Memor

y

Memor

y

VENTILA

TOR P

ARAM

ETERS

FLO

W/

VO

L

FLO

W/

VO

L

TREND

TEST

ZER

O

SENSORS

MORE

PRESSURE

PRESSURE

PRESSURE

TIM

ING

VOLUM

E

FLO

W

W

AVEFORM

SCREENS

OTHER MEASUREMENTS

SPECIAL FUNCTIONS

%RH

O

2

_ , @

_ , @

11

22

AB

C

AB

C

55

JK

L

JK

L

44

GH

I

GH

I

66

MN

O

MN

O

88

00

TU

V

TU

V

77

PQ

RS

PQ

RS

99

WX

YZ

WX

YZ

33

DE

F

DE

F

GAS FLOW ANALYZER

GAS FLOW ANALYZER

VT

M

OB

ILE

VT

M

OB

ILE

Low

Pressure

Adapter

edg08f.eps

Figure 5-1. Measuring Low Pressure

Advertising