Fluke Biomedical VT Mobile User Manual

Page 46

Advertising
background image

VT MOBILE

Operators Manual

4-2

5. Using parts from the Accessory Kit and the connectors provided with the sensors,

make connections between the ventilator and the ACCU LUNG as shown in Figure

4-1. Use the sequence: ventilator Y connector Æ High-Flow Sensor (blue stripe

toward the ACCU LUNG) Æ Temperature/RH Sensor Æ Oxygen Sensor Æ ACCU

LUNG. Vertically align all sensors.

6. Set up the ventilator for a characteristic breath pattern. For example, you could set 10

BPM at 7.5 lpm.

7. Set the ACCU LUNG for C20 compliance (both outer springs engaged) and Rp50

resistance (positioned as shown in Figure 4-1.)

Fluke Biomedical
ACCU LUNG

High Flow Sensor

Oxygen

Sensor

To

Ventilator

Temp./RH Sensor

ede06f.eps

Figure 4-1. Ventilator Connections

8. Apply power to all elements of the test setup. For the Analyzer, press H on, wait for

the Zeroing screen, and then press

G

. The Analyzer is ready to use once the Tidal

Volume screen appears.

9. Press the access keys shown in Table 4-1 to display the various ventilator parameters.

Once you have pressed the first key, you can continue pressing that key to view other

parameters or press

G

to view all the parameters.

10. As described in Table 4-1, the Analyzer calculates 16 breath parameters in Local

Mode. Note that you can view all 16 parameters and other information on one screen

in Remote mode when using the VT for Windows software.

11. Figure 4-2 shows the ventilator parameter screens.
12. Figure 4-3 shows additional oxygen, temperature, relative humidity, and barometric

pressure parameter screens.

Advertising