Gas settings – Fluke Biomedical VT Mobile User Manual

Page 93

Advertising
background image

Appendices

VT for Windows

A

A-27

Gas Settings

The Analyzer can compensate for the effects of gas temperature, viscosity, molecular

weight, and relative humidity on flow measurement, as well as correct flow calculations

to one of several standard conditions. The Gas Settings option gives you access to these

settings.
To change the gas settings, select Gas Settings from the Setup menu. The dialog box

shown in Figure A-34 appears, allowing you to make the changes.

edg74s.bmp

Figure A-34. Gas Settings Dialog Box

When you select the Advanced option, the Gas Settings dialog box lets you adjust the

temperature, viscosity, molecular weight, and relative humidity for both the inhaled and

exhaled gases. In addition, you can adjust the ambient temperature and have the flow

adjusted to one of four standard correction modes (ATP, BTPS, STPD0, and STPD21).

See Figure A-35.

Advertising