Measuring airway pressure – Fluke Biomedical VT Mobile User Manual
Page 53
Advertising
![background image](/files/785355/content/doc053.png)
Measuring Pressure
Measuring Airway Pressure
5
5-5
Measuring Airway Pressure
The Analyzer uses the same pressure transducer for low pressure and airway pressure
measurements. It derives airway pressure internally through the connections for high- or
low-flow measurements. Consequently, airway pressure measurements are available with
the High- or Low-Flow Sensor connected during ventilator parameter checks.
Figure 5-5 illustrates ways to view the airway pressure.
OTHER
MEASUREMENTS
WAVEFORM
SCREENS
edg05f.eps
Figure 5-5. Accessing Airway Pressure Measurements
Advertising