Using the VT MOBILE

Making a Measurement More Accurate

3

3-17

Oxygen Sensor

Source

edg20f.eps

Figure 3-4. Oxygen Sensor Calibration

1. Press P to display the Oxygen measurement screen.
2. Press G to access the Oxygen calibration screen.

ede28s.bmp

3. Connect the High-Flow Sensor and then apply dry air (20.9 % oxygen) at 10 lpm to

either side of the sensor.

4. Press F twice to access the Calibration screen. Wait for the two-minute timer to

count down.

5. Apply 100 % oxygen at 10 lpm.
6. Press G twice and then press F once to access the Calibration screen.
7. After the two-minute interval, the Oxygen measurement screen appears and the

calibration is complete.

Using the Temperature and Relative Humidity Sensor

Use the optional Temperature and Relative Humidity Sensor to compensate for the

conditions of the gas flowing through the High- or Low-Flow Sensor. The Analyzer uses

the gas temperature to improve the accuracy of the flow measurements. Relative

humidity readings provide a reference point for some correction factors. Connect this

sensor to the Analyzer as shown in Figure 3-5.