Playback controls – Fluke Biomedical VT Mobile User Manual

Page 84

Advertising
background image

VT MOBILE
Operators Manual

A-18

Playback Controls

Use the controls shown in Table A-3 as aids in data review and recording.

Table A-3. Playback Controls

Button Name

Description

Record

The Record button initiates writing data to the open data
file. The software places a Time event marker in the
data file, indicating the beginning of recording.

Play

The Play button initiates data playback from an open
file.

Stop

The Stop button terminates data recording or playback.
The software leaves the data file open.

Pause

The Pause button stops data playback temporarily.
Pause files before printing so that the software can
capture the plot for printing.

Start

This button rewinds the data file to its beginning. This
only affects data playback. The software performs all
recording at the end of the data file.

Rewind

This button backs up the data file during playback to the
previous event marker in the file. Automatic event
markers provide periodic events to which you can
rewind or forward the file.

Continuous
Play On

Enable this button during data playback to allow for
continuous play. In continuous play, the data file
automatically rewinds to the beginning as soon as the
end of the file occurs. The button appears red when
enabled and black when disabled.

Fast
Forward

This button advances the data file during playback to
the next event marker in the file. Automatic event
markers provide periodic events that enable rewinding
or forwarding of the file.

End

This button advances the data file to its end. This only
affects data playback.

Speed
Control

The speed control adjusts the playback speed. The
number shown is the time in milliseconds (ms) between
waveform samples. The data sample rate for the
Analyzer is 20 ms. Therefore, a number larger than 20
results in slower-than-normal playback; smaller
numbers yield a fast playback speed.