B. specifications, General specifications, Battery power supply – Fluke Biomedical VT Mobile User Manual

Page 99: External power supply, Measurement specifications, Low-pressure port, Specifications, Appendix b

Advertising
background image

B-1


Appendix B

Specifications

General Specifications

Size:

8 L x 1.5 H x 4 W inches (20 L x 3.8 H x 10 W cm)

Weight:

1 lb (0.45 kg)

Temperature:

Operating: 10 to 40

°

C (50 to 104

°

F)

Storage: -25 to 50

°

C (-13 to 122

°

F)

Maximum Humidity, Operating: 80 % relative humidity up to 31

°

C (88

°

F), decreasing linearly to 50 %

relative humidity at 40

°

C (104

°

F).

Maximum Humidity, Storage:

95 %

Barometric:

7 to 18 psia, operating; -1000 to 10000 ft (787.9 – 522.7 mmHg), storage

Battery Power Supply

Voltage:

9 VDC

Power Consumption:

< 70 mA

Battery Life:

> 7 hours

External Power Supply

Output Voltage:

12 – 15 V

Output Current:

1.2 A

Measurement Specifications

Low-Pressure Port

Maximum Applied Pressure:

5 psi

Operating Pressure (Differential):

-20 to 120 cmH

2

O

Operating Pressure (Common Mode):

NA

Span Accuracy:

± 2 % of reading or 1.5 mmHg, whichever is greater

Frequency Response:

> 10 Hz

Resolution:

0.1 mmHg

Sample Rate:

100 Hz

Fittings:

Flow Connector with 2 tubes "T" connected to a
single fitting

Note: Apply no fluid to the port.