Trend test screens, Zero sensors screens, More key screens – Fluke Biomedical VT Mobile User Manual

Page 35

Advertising
background image

Using the VT MOBILE

Screens

3

3-11

Trend Test Screens

Press Q to access the Trend Test screens, which allow you to record instances (or

events) of a selected ventilator parameter falling outside limits that you establish. The

opening screen in the trend test setup sequence appears below:

ede40s.bmp

Zero Sensors Screens

Press R to access the Zero Sensors screens, which allow you to clear the offset

resulting from drift inherent in the differential pressure transducer. The opening screen in

the zero sensors sequence appears below:

ede41s.bmp

This screen also appears automatically at a set interval to remind you to zero the sensors

if Zero Warn is set to ON in the setup procedure.

More Key Screens

Barometric Pressure Screen

Press S Æ E to access the opening Barometric Pressure Screen, which allows

you to view or change the barometric pressure that is automatically measured by an

internal sensor. Under normal operating conditions, barometric pressure should not need

adjustment. The following is a sample of the opening Barometric Pressure Screen:

ede31s.bmp

Battery Screen

Press S Æ C Æ E to access the Battery Screen, which allows you to view the

battery level. A sample of the 9-volt battery reading (when the battery eliminator is not

connected) appears below:

ede43s.bmp

If the battery eliminator is connected, this screen shows approximately 12 volts.

Advertising