Vt mobile, Statistics, Zeroing – Fluke Biomedical VT Mobile User Manual

Page 72: Setup

Advertising
background image

VT MOBILE
Operators Manual

A-6

VT Mobile

This selection allows you to download files from the Analyzer.

Statistics

The Statistics menu gives access to the clear function for any or all statistics. The Clear

Statistics function clears the statistics for all the signals and the breath parameters at

once. Clear individual signal statistics by selecting the appropriate option. Refer to Figure

A-6.

edg84s.bmp

Figure A-6. Statistics Menu

Zeroing

The Zeroing menu (Figure A-7) allows you to zero all of the signals manually. The Zero

All function zeros all the signals except oxygen. The Oxygen function allows you to zero

sensor drift at 21% oxygen and calibrate at 100% oxygen.

edg85s.bmp

Figure A-7. Zeroing Menu

Setup

The Setup menu (Figure A-8) provides access to some of the user setup features found in

the Analyzer, including:

Breath Detect:

Use the Breath Detect menu to switch breath detection modes and set the breath

detect threshold for the current flow port.

Gas Settings:

Advertising