Conducting a Leak Test

Overview

The Leak test is for measuring the leak rate of a sealed vessel or test lung over a period of

time.

Starting a Leak Test

Start the Leak test by selecting the Leak Test tab (accessed by selecting File Æ View File

Controls). Then select Start Leak Test. (You can also start a Leak test by selecting Leak

Test from a pressure Plot menu on the Monitor Plots or Single Plot screen.) The dialog

box shown in Figure A-17 now appears, giving you the chance to choose the pressure

channel on which you want to run the test. In addition, you can change the displayed

pressure units. If you choose to enter a compliance value for the system under test, the

software calculates the volume lost during the test. Enter the test time to set the length of

time that the Leak test runs. The Leak test automatically stops at the end of that amount

of time. Click the OK button to begin the Leak test.

Running a Leak Test

The Leak test plots the chosen pressure channel (airway, low, or high pressure) over the

chosen length of time. The Leak test also displays several values calculated during the

test. Refer to Figure A-18.

Current Pressure

This is the pressure currently measured.

Initial Pressure