Starting the software – Fluke Biomedical VT Mobile User Manual

Page 68

Advertising
background image

VT MOBILE
Operators Manual

A-2

Starting the Software

With the software installed and the Analyzer connected, you are ready to start the

software application. Upon first use, the application attempts to establish communications

with the Analyzer.
The software checks the serial communication ports (also called Comm Ports) listed in

the Windows Registry. If the software cannot find the Analyzer, the screen in Figure A-1

appears, allowing you to set the communication port used for connecting the Analyzer.

edg53s.bmp

Figure A-1. Comm Port Selection

Once the software has established communication with the Analyzer, it creates an

initialization file to read during future use. The software uses the information in the

initialization file to know where to check for the Analyzer.
An error can occur if the communication port is in use by another device and/or software

application. To solve this error, close the application that is using the required

communication port.
Finally, if you need to manually set or change the communication port for the Analyzer,

start VT for Windows and select Communications Æ Comm Port from the pull-down

menu. Select the appropriate communication port. You can tell that VT Mobile has

established communications when it starts displaying waveforms and changing values for

flow and pressure.