Cleaning, Verifying oxygen sensor calibration – Fluke Biomedical VT Mobile User Manual

Page 62

Advertising
background image

VT MOBILE

Operators Manual

7-2

1

2

ede09f.eps

Figure 7-1. Changing the Battery

Cleaning

Clean the Analyzer with a damp cloth on the exterior only. Use a solution of 70%

isopropyl alcohol. Fluke Biomedical does not recommend other solvents.

Verifying Oxygen Sensor Calibration

Verify the oxygen sensor calibration at the beginning of any day during which you expect

to measure oxygen concentrations. Use the following verification procedure:
1. Connect the Oxygen Sensor as shown in Figure 7-2.
2. Introduce a 10 lpm or greater flow of 100% oxygen through the Oxygen Sensor “T”.
3. Press P to access the O2 display.
4. Ensure that the O2 screen oxygen concentration is 100 % ± 2 % of full scale. If the

concentration is less than 98 %, perform the oxygen sensor calibration as described

below.

5. Replace the oxygen sensor after one year (or longer if no replacement message

appears during calibration.)

Note

The Oxygen Sensor is not calibrated at the factory.