Measurement parameter specifications, Display screen, Operational modes – Fluke Biomedical VT Mobile User Manual

Page 103: Gas types, Serial communications specifications

Advertising
background image

Appendices

Specifications

B

B-5

Peak Inspiratory Pressure (PIP)

0.1 cmH

2

O

± 120 cmH

2

O

± 3 % or 2 cmH

2

O

Inspiratory Pause Pressure (IPP)

0.1 cmH

2

O

± 120 cmH

2

O

± 3 % or 2 cmH

2

O

Mean Airway Pressure (MAP)

0.1 cmH

2

O

± 80 cmH

2

O

± 3 % or 1 cmH

2

O

Positive-End Expiratory Pressure (PEEP)

0.1 cmH

2

O

- 5 to 40 cmH

2

O ± 3 % or 1 cmH

2

O

Peak Expiratory Flow (PEF)

0.01 lpm

0 to 150 lpm

± 3 % or 1 lpm

Peak Inspiratory Flow (PIF)

0.01 lpm

0 to 150 lpm

± 3 % or 2 lpm

Inspiratory Time (Ti)

0.01 s

0 to 60 s

0.5 % or 0.02 s

Expiratory Time (Te)

0.01 s

0 to 90 s

0.5 % or 0.01 s

Display Screen

64 x 128 pixels, reflective LCD, blue on yellow

Operational Modes

Standalone
Controlled by VT for Windows PC Software (version 2.01.00 or higher)

Gas Types

Air, O

2

, Heliox, CO

2,

N

2

, N

2

O, N

2

O bal O

2

, He bal O

2

, N

2

bal O

2

User

Serial Communications Specifications

4-pin miniature RS-232 serial port located on upper-left side of the Analyzer
RS232 compatible with the VT for Windows software application (version 2.01.00 or

higher.)

Measurement Parameter Specifications

Parameter Resolution

Range

Accuracy

Inspiratory and Expiratory Tidal Volume (VT)

0.1 ml

> 10 l

± 3 %

Expiratory Minute Volume (MV)

0.001 lpm

0 to 60 l

± 3 %

Breath Rate (BPM)

0.1 bpm

2 to 150 bpm

± 1 %

Inspiratory, Expiratory Time (I:E)

0.01 s

0.25 to 9.99 s

± 2 % or 0.1 s

Advertising