High-flow port, Low-flow port – Fluke Biomedical VT Mobile User Manual

Page 101

Advertising
background image

Appendices

Specifications

B

B-3

High-Flow Port

Maximum Flow Rate (absolute value):

200 lpm

Operating Flow Range:

± 150 lpm

Accuracy:

± 3 % of reading or ± 2 % of range, whichever is
greater

Floor for Absolute Accuracy:

25 lpm

Resolution:

0.01 lpm

Frequency Response:

> 25 Hz or t

10-90

< 40 ms

Sample Rate:

100 Hz

Dynamic Resistance:

< 2 cmH

2

O @ 60 lpm

Low-Flow Dropout:

2.5 lpm

Breath-Detect Threshold:

4 lpm (user selectable)

Volume Range:

> ± 60 l

Tidal Volume Accuracy:

± 3 % or reading or ± 20 ml, whichever is greater

Low-Flow Port

Maximum Flow Rate (absolute value):

35 lpm

Operating Flow Range:

± 25 lpm

Accuracy:

± 3 % of reading ± 1% of range, whichever is greater

Floor for Absolute Accuracy:

3 lpm

Resolution:

0.01 lpm flow > 1 lpm

Freq. Response:

> 25 Hz or t

10-90

< 40 ms

Sample Rate:

100 Hz

Dynamic Resistance:

< 2.5 cmH

2

O @ 5 lpm

Low-Flow Dropout:

0.24 lpm

Breath-Detect Threshold:

1 lpm (user selectable)

Volume Range:

± 60 l

Tidal Volume Accuracy:

± 3 % of reading or ± 5 ml