Using the VT MOBILE

Making a Measurement More Accurate

3

3-15

The Analyzer requires you to zero immediately after power on. If Zero Warn is ON, the

Analyzer prompts you for an optional zero after the first five minutes and then after each

subsequent 30-minute interval. To zero at the prompt, follow these steps:
1. Place the Analyzer on a flat surface with sensors/adapters disconnected at the

Analyzer (Figure 3-2.)

edg17f.eps

Figure 3-2. Flow Sensors and Pressure Adapters Disconnected



2. Press G. Press R at any time to initiate an unprompted zero. Generally, it is

appropriate to zero before any pressure or flow measurement.

3. Reconnect the sensors/adapters (Figure 3-3.)

edg18f.eps

Figure 3-3. Flow Sensors and Pressure Adapters Connected