Using the VT MOBILE

Using Trend Test

3

3-13

8. When both Low and High limit values are correct, press E again to advance to

the Start Trend Test screen.

ede46s.bmp


9. Before starting the test, ensure that the ventilator is on, has reached steady-state

conditions, and that the displayed breath parameter is as expected. Press G (Start)

to begin the test.

10. The Analyzer displays the Trend Test screen while the trend test is in progress. The

trend test screen shows the parameter being tested, the current value, the upper and

lower limits, and the number of events (readings falling outside the limits.)

ede47s.bmp


11. To terminate the test, press G (Stop). The Trend Summary screen appears. If there

are no events, the Trend Summary screen is not available.

ede48s.bmp

ede49s.bmp


12. Alternatively, press G (More) to display related parameters.

ede50s.bmp