Correction modes, Flow zeroing – Fluke Biomedical VT Mobile User Manual

Page 56

Advertising
background image

VT MOBILE

Operators Manual

6-2

Correction Modes

The Analyzer obtains the ambient pressure from the measured barometric pressure. The

optional Temperature and Relative Humidity Sensor supplies the measured gas

temperature and relative humidity. Table 6-1 describes correction modes that are

dependent on these measurements.

Table 6-1. Correction Modes

Mode Name

Description

ATP Ambient

Temperature

and Pressure

The Analyzer corrects the flow rate (and therefore, volume) such that the
reported flow equals the flow if the gas temperature and pressure are at
ambient (room) temperature and pressure.

BTPS Body Temperature and

Pressure Saturated
with water vapor

The Analyzer corrects the flow rate (and therefore, volume) such that the
reported flow equals the flow if the gas temperature and pressure were
changed to body conditions (37

°

C and ambient pressure). Further, the

Analyzer compensates the flow to add the amount of water vapor as if to
make the gas fully saturated (i.e. 100% relative humidity). In this
correction, the Analyzer uses the relative humidity reading to determine
the amount of water vapor that would need to be added.

STPD

0

Standard

Temperature

and Pressure Dry,
0

°

C

The Analyzer corrects the flow rate (and therefore, volume) such that the
reported flow equals the flow if the gas temperature and pressure were
changed to standard conditions (0

°

C and 760 mmHg). Further, the

Analyzer compensates the flow to subtract the amount of water vapor as
if the gas was dry (i.e. 0% relative humidity). In this correction, the
Analyzer uses the relative humidity reading to determine the amount of
water vapor that would need to be subtracted.

STPD

21

Standard Temperature

and Pressure Dry,
21

°

C

The Analyzer corrects the flow rate (and therefore, volume) such that the
reported flow equals the flow if the gas temperature and pressure were
changed to standard conditions (21

°

C and 760 mmHg). Further, the

Analyzer compensates the flow to subtract the amount of water vapor as
if the gas was dry (i.e. 0% relative humidity). In this correction, the
Analyzer uses the relative humidity reading to determine the amount of
water vapor that would need to be subtracted.

Flow Zeroing

The zero function can be set to prompt you automatically at pre-determined intervals or

only when you request the function by pressing R. The Analyzer Zero Warn is set to

ON at the factory to remind you to zero out any offset in the Flow measurements. The

Zero Warn screen reminds you at power on, after the first five minutes, and then after

each 30-minute period. After the power-on zero warning, you can choose to Zero or

Cancel when the reminder appears; either action returns the Analyzer to the same

function.

W

Caution

Setting Zero Warn to OFF may result in introduction of offset

into the Flow and Volume measurements and cause errors in

the assessment of those measurements against the medical

device under test specification. Fluke Biomedical strongly

recommends that you leave Zero Warn ON.

The differential pressure transducer used to measure the flow rate is subject to baseline

(offset) drift. You will see this most pronounced during the Analyzer warm-up period.

For maximum accuracy, observe a 40-minute warm-up, followed by the zeroing

procedure.

Advertising